Technical Data

Infrared thermometer

GIS800‑16

Article number

‎3 601 K83 B..‎‎‎‎

Working range

0.1−5 m

Measuring rangeA)

−40 °C to +800 °C

Temperature resolution

0.1 °C

Optics (measuring distance : measured surface ratio)B)C)

20 : 1

Size of display

2.4"

Laser class

2

Laser type

< 1 mW, 640−660 nm

Divergence of the laser beam (full angle)

< 1.5 mrad

Max. altitude

2000 m

Pollution degree according to IEC 61010-1

2D)

Relative air humidity max.

90 %

Power supply

  • Rechargeable battery (Li-ion)

10.8 V/12 V

  • Non-rechargeable batteries (alkaline manganese, with battery adapter)

4 × 1.5 V LR6 (AA)

  • Rechargeable batteries (NiMH, with battery adapter)

4 × 1.2 V HR6 (AA)

Operating time

  • Rechargeable battery (Li-ion)E)F)

18 h

  • Non-rechargeable batteries (alkaline manganese)

12 h

WeightG)

0.36 kg

Dimensions (length × width × height)

119 × 73 × 212 mm

Protection ratingH)

IP54

Recommended ambient temperature during charging

0 °C to +35 °C

Permitted ambient temperature during operation

–10 °C to +50 °C

Permitted ambient temperature during storage without a rechargeable battery

–20 °C to +70 °C

Permitted ambient temperature during storage with a rechargeable battery

–20 °C to +50 °C

Recommended rechargeable batteries

GBA 10.8V…
GBA 12V…

Recommended battery chargers

GAL 12…
GAX 18…

A)

Maximum measuring range of the measuring tool – with contact temperature measurements, the temperature probe in use can have a smaller measuring range.

B)

In accordance with standard VDI 5585 (average)

C)

Refers to infrared measurement, see figure:

D)

Only non-conductive deposits occur, whereby occasional temporary conductivity caused by condensation is expected.

E)

Depending on battery in use

F)

At an ambient temperature of 20–30 °C

G)

Weight without lithium-ion rechargeable battery/battery adapter/non-rechargeable batteries (you can find the weight of the lithium-ion battery at www.bosch-professional.com.)

H)

excluding lithium-ion battery pack/non-rechargeable batteries/rechargeable batteries, in the vertical position

The serial number (9) on the type plate is used to clearly identify your measuring tool.

Measuring accuracy

At measured value

At aperture

At measuring distance

Measuring accuracy

Surface temperatureA)

−40 °C to −30.1 °C

50 mm

10 cm to 50 cm

±5.0 °C

−30 °C to −20.1 °C

57 mm

10 cm to 50 cm

±4.5 °C

−20 °C to −10.1 °C

57 mm

30 cm to 100 cm

±3.5 °C

−10 °C to 0 °C

152 mm

30 cm to 100 cm

±2.5 °C

+0.1 °C to +100 °C

152 mm

30 cm to 100 cm

±1.5 °C

+100.1 °C to +500 °C

152 mm

30 cm to 100 cm

±1.5 %

+500.1 °C to +800 °C

50 mm

10 cm to 50 cm

±1.5 %

Contact temperature (only with type K temperature probe)B)

−40 °C to +333 °C

±2.5 °C

+333.1 °C to +400 °C

±0.75 %

A)

At an ambient temperature of +21 °C to +25 °C, an emissivity of ≥ 0.95, with worklight and laser switched off; plus usage-dependent deviation (e.g. reflection)

B)

according to IEC EN 60584‑1: Type K, Class 2